Uploaded on 2024/05/28 OPERATION LOG OF THE RESOLVE INSTRUMENT Ordered by UT-Date of Occurrence (old to new). -------------------------------- 2024/01/21 ~ -------------------------------- Comment: - Regular operation procedure for gate valve closing started with combination of 55Fe intermittent operation. Observation(s) related to this note: - 000154000 / Perseus_CAL1 -------------------------------- 2024/02/18 13:21:49 -------------------------------- Comment: - Unusual telemetry output. PSP Syslog error (PSPB0) was reported during OBSID 300005010 (Kepler_SNR). It was detected during pointing to night earth. There is no impact on astronomical observations. Observation(s) related to this note: - 300005010 / Kepler_SNR -------------------------------- 2024/02/18 - 2024/02/20 -------------------------------- Comment: - Resolve calibration observation; FWE operation. Observation(s) related to this note: - 300080010 / SN1006 - 100010010 / 1E0102.2-7219 -------------------------------- 2024/02/28 23:27:00 -------------------------------- Comment: - Unusual telemetry output. PSP Syslog error (PSPA0) was reported. It was detected during pointing to OBSID 300044010 (Galactic_Center_1). We are investigating what is happening near that time of error. Observation(s) related to this note: - 300044010 / Galactic_Center_1 -------------------------------- 2024/03/06 - 2024/03/10 -------------------------------- Comment: - Resolve calibration observation; FWE and PSP operations. Observation(s) related to this note: - 100011010 / HR1099 -------------------------------- 2024/03/10 ~ -------------------------------- Comment: - Regular operation procedure updated. Add one Fe55 illumination starting at ~9hr after an attitude maneuver ends. Observation(s) related to this note: - 300072010 / PDS456 -------------------------------- 2024/03/10 23:36:00 -------------------------------- Comment: - Switch from PSP forced midres to PSP nominal mode. Observation(s) related to this note: - 300072010 / PDS456 -------------------------------- 2024/03/15 01:07:48 -------------------------------- Comment: - Unusual telemetry output. PSP Syslog error (PSPB1) was reported. It was detected during pointing to OBSID 300072010 (PDS456). We are investigating what is happening near that time of error. Observation(s) related to this note: - 300072010 / PDS456 -------------------------------- 2024/03/24 06:03:00 -------------------------------- Comment: - Regular operation procedure updated. Change one Fe55 illumination starting at ~9hr after an attitude maneuver ends to at ~9.36hr after an attitude maneuver ends. Observation(s) related to this note: - 300065010 / Cyg_X-3 -------------------------------- 2024/04/18 12:40:03 -------------------------------- Comment: - FWE memory dump: FWE memory was dumped for regular health-check during pointing to OBSID 100008010 (Cygnus_Loop). Data for scientific purposes is not affected. Observation(s) related to this note: - 100008010 / Cygnus_Loop -------------------------------- 2024/04/30 16:46:12 -------------------------------- Comment: - ADRC Recycle with S3: Commanded ADR Recycle was executed by running S3 for regular health-check during pointing to OBSID 300055010 (W49B_East). Data for scientific purposes is not affected. Observation(s) related to this note: - 300055010 / W49B_East -------------------------------- 2024/05/07 19:56:35 -------------------------------- Comment: - Unusual telemetry output: PSP Syslog error (PSPA0) was reported. It was detected during pointing to OBSID 300056010 (W49B_West). We are investigating what is happening near that time of error. Observation(s) related to this note: - 300056010 / W49B_West